Experimental design techniques for optimizing psychrometric humidity sensor calibration

A. Frattolillo
1999-01-01

Abstract

This paper tackles the design of a calibration using experimental design techniques. In particular, the optimum calibration pian for psychrometers, treated as a complex measurement chain, is identified by suitably elaborating the errar propagation law suggested by the ISO Guide. The obtained theoretical results are verified on a number of real calibrations, carried out in an humidity European Accredited Laboratory. The main advantages of the proposed approach consist in i) calibrati an plans with a reduced number of calibration points; ii) calibrati an curve applicable to to the whole temperature range; iii) linear rather than complex regression technique always usable.
1999
sensor calibration, humidity sensors, psychrometer, experimental design.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Questionario e social

Condividi su:
Impostazioni cookie