Advancements towards the implementation of clinical phase-contrast breast computed tomography at Elettra

Golosio B.;Fanti V.;
2019-01-01

Abstract

Breast computed tomography (BCT) is an emerging application of X-ray tomography in radiological practice. A few clinical prototypes are under evaluation in hospitals and new systems are under development aiming at improving spatial and contrast resolution and reducing delivered dose. At the same time, synchrotron-radiation phase-contrast mammography has been demonstrated to offer substantial advantages when compared with conventional mammography. At Elettra, the Italian synchrotron radiation facility, a clinical program of phase-contrast BCT based on the free-space propagation approach is under development. In this paper, full-volume breast samples imaged with a beam energy of 32 keV delivering a mean glandular dose of 5 mGy are presented. The whole acquisition setup mimics a clinical study in order to evaluate its feasibility in terms of acquisition time and image quality. Acquisitions are performed using a high-resolution CdTe photon-counting detector and the projection data are processed via a phase-retrieval algorithm. Tomographic reconstructions are compared with conventional mammographic images acquired prior to surgery and with histologic examinations. Results indicate that BCT with monochromatic beam and free-space propagation phase-contrast imaging provide relevant three-dimensional insights of breast morphology at clinically acceptable doses and scan times.
2019
Inglese
26
4
1343
1353
11
http://journals.iucr.org/s/
https://onlinelibrary.wiley.com/iucr/doi/10.1107/S1600577519005502
Esperti anonimi
internazionale
scientifica
breast CT; free space propagation; phase contrast; single photon counting
Longo, R.; Tromba, G.; Zanconati, F.; Tonutti, M.; Taibi, A.; Russo, P.; Rigon, L.; Sarno, A.; Pacile, S.; Oliva, P.; Mettivier, G.; Golosio, B.; Fedo ...espandi
1.1 Articolo in rivista
info:eu-repo/semantics/article
1 Contributo su Rivista::1.1 Articolo in rivista
262
25
partially_open
File in questo prodotto:
File Dimensione Formato  
JSR_2019_S1600577519005502.pdf

Solo gestori archivio

Tipologia: versione editoriale
Dimensione 3.34 MB
Formato Adobe PDF
3.34 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
2944840_Longo_et_al-2019-JSR.pdf

accesso aperto

Tipologia: versione pre-print
Dimensione 3.32 MB
Formato Adobe PDF
3.32 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Questionario e social

Condividi su:
Impostazioni cookie