Ocular perfusion pressure control during pars plana vitrectomy: testing a novel device

Querzoli, Giorgio;
2017-01-01

Abstract

Purpose: To study the efficacy of a novel device intended to control infusion pressure based on mean ocular perfusion pressure (MOPP) during pars plana vitrectomy (PPV). Methods: An arm blood pressure cuff connected to a vitrectomy machine calculated mean arterial pressure (MAP), while a pressure sensor close to the infusion trocar measured intraocular pressure (IOP). MOPP was calculated in real time in 36 consecutive patients undergoing PPV, who were divided into two groups. The device lowered IOP every time that calculated MOPP fell below 30 mmHg in the Control ON group (18 patients), while no action was taken in the Control OFF group (18 patients). Results: Baseline IOP and blood pressure were similar between groups. The Control ON group had significantly lower average intraoperative IOP (30.5 ± 2.1 vs. 35.9 ± 6.9 mmHg; p = 0.002) and higher MOPP (56.4 ± 5.9 vs. 49.7 ± 6.1 mmHg) than the Control OFF group. The Control ON group also spent less time at MOPP <â 10â mmhg and <â 30â mmhg: 0 vs. 3.40â â±â 2.38â min (pâ <â 0.001) 9.91â â±â 7.15 16.13â â±â 8.12â min (pâ =" 0.02)," respectively. conclusions: the mopp control device effectively maintained lower iop higher throughout surgery. it also helped avoid dangerous peaks dips, allowing patients to spend less time at of <â 10 <â 30â mmhg. < div>
2017
2017
Inglese
255
12
2325
2330
6
link.springer.de/link/service/journals/00417/index.htm
Esperti anonimi
scientifica
Intraocular pressure; Mean arterial pressure; Ocular perfusion pressure; Pars plana vitrectomy; Blood Pressure; Female; Glaucoma; Humans; Intraocular Pressure; Male; Middle Aged; Monitoring, Intraoperative; Perfusion; Reproducibility of Results; Vitrectomy; Ophthalmology; Sensory Systems; Cellular and Molecular Neuroscience
no
Rossi, Tommaso; Querzoli, Giorgio; Gelso, Aldo; Angelini, Giampiero; Rossi, Alessandro; Corazza, Paolo; Landi, Laura; Telani, Serena; Ripandelli, Guid ...espandi
1.1 Articolo in rivista
info:eu-repo/semantics/article
1 Contributo su Rivista::1.1 Articolo in rivista
262
9
reserved
File in questo prodotto:
File Dimensione Formato  
2017 GACE - Ocular perfusion pressure control during pars plana vitrectomy; testing a novel device.pdf

Solo gestori archivio

Tipologia: versione editoriale
Dimensione 775.16 kB
Formato Adobe PDF
775.16 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11584/234239
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 6
social impact

Questionario e social

Condividi su:
Impostazioni cookie