Equipment testing and tuning: the cold-knife cryomicrotome microm HM-560

FERRI, GIAN LUCA;COCCO, CRISTINA;
2002-01-01

Abstract

A major innovative feature of the Microm HM-560 cryomicrotome is the independent control of specimen and knife temperatures. We used such equipment with a variety of tissues, and fixation and freezing procedures. High-quality sectioning was reproducibly obtained using 1) a low temperature setting for the sectioning blade ("cold knife," about -33°C); 2) a comparatively high temperature for the specimen, and 3) a suitable mounting medium, which would remain solid up to about 0°C. Specimen temperature was set between -8° and -15°C for 4-μm sectioning, higher temperatures (-1° to -8°C) being appropriate when cutting at 10 to 20 μm. Under such conditions, disposable blades were effective throughout, while a modified antiroll plate profile further enhanced usability. After intensive use for almost 3 years, by more than 15 different users, the cryomicrotome is in excellent working order.
2002
10
381
386
6
Esperti anonimi
Impact Factor 1.54
Ferri, GIAN LUCA; Cocco, Cristina; Melis, Gv; Aste, L.
1.1 Articolo in rivista
info:eu-repo/semantics/article
1 Contributo su Rivista::1.1 Articolo in rivista
262
4
none
Files in This Item:
There are no files associated with this item.