EMC Characterization of Implantable Cardiac Medical Devices in an Anechoic Chamber
SPANO, IVAN LUIGI;SULIS, SARA;SERPI, ALESSANDRO;MARONGIU, IGNAZIO;GATTO, GIANLUCA
2014-01-01
Abstract
A novel electromagnetic characterization of Implantable Cardiac Medical Devices (ICMDs) is presented in this paper. It has been developed in order to determine the immunity level of ICMDs against Radio-Frequency (RF) interference and consists of several Sensitivity Tests, which have been performed in accordance with EMC international standards. The tests have been carried out in an anechoic RF chamber, which guarantees an appropriate management of noise signal injection, shielding against external RF noise sources at the same time. In this way, the operating thresholds of an Implantable Cardiac Defibrillator (ICD) have been achieved at different heartbeat frequency, with and without high-frequency noise. As a result, the comparison between the operating thresholds highlight the electromagnetic susceptibility of the ICD under test.File | Size | Format | |
---|---|---|---|
P41#2014#EMC-EU#IEEE_Paper.pdf Solo gestori archivio
Description: EMC Characterization of Implantable Cardiac Medical Devices in an Anechoic Chamber
Type: versione editoriale
Size 1.78 MB
Format Adobe PDF
|
1.78 MB | Adobe PDF | & nbsp; View / Open Request a copy |
P41#2014#EMC-EU#PostPrint.pdf Solo gestori archivio
Description: EMC Characterization of Implantable Cardiac Medical Devices in an Anechoic Chamber (Post Print)
Type: versione post-print
Size 1.76 MB
Format Adobe PDF
|
1.76 MB | Adobe PDF | & nbsp; View / Open Request a copy |