Experimental design techniques for optimizing psychrometric humidity sensor calibration

A. Frattolillo
1999-01-01

Abstract

This paper tackles the design of a calibration using experimental design techniques. In particular, the optimum calibration pian for psychrometers, treated as a complex measurement chain, is identified by suitably elaborating the errar propagation law suggested by the ISO Guide. The obtained theoretical results are verified on a number of real calibrations, carried out in an humidity European Accredited Laboratory. The main advantages of the proposed approach consist in i) calibrati an plans with a reduced number of calibration points; ii) calibrati an curve applicable to to the whole temperature range; iii) linear rather than complex regression technique always usable.
1999
sensor calibration, humidity sensors, psychrometer, experimental design.
Files in This Item:
There are no files associated with this item.