An Automatic Sensing Test Procedure for Implantable Cardioverter Defibrillator

SPANO, IVAN LUIGI;SULIS, SARA;SERPI, ALESSANDRO;MARONGIU, IGNAZIO;GATTO, GIANLUCA;
2015-01-01

Abstract

An Automatic Sensing Test procedure (AST) for evaluating the sensing performances of Implantable Cardioverter Defibrillators (ICDs) is presented in this paper. AST is based on the Sensitivity Test described by the Standard CEI EN 45502-2-1 for pacemakers and consists of determining the ICD sensing thresholds at different heartbeat frequencies. Consequently, an extensive evaluation of ICD sensing performances can be achieved. AST has been implemented by means of virtual instruments developed in the Labview environment and has been validated experimentally through several tests. These have been performed inside an anechoic RF chamber. Such an environment guarantees proper shielding against external electromagnetic interferences, thus ensuring reliability and repeatability of the tests. The discussion of the results highlights the usefulness of the proposed procedure.
2015
Inglese
2015 IEEE International Symposium on Medical Measurements and Applications (MeMeA) Proceedings
978-1-4799-6477-2
186
190
5
2015 IEEE International Symposium on Medical Measurements and Applications, MeMeA 2015
Esperti anonimi
May 7-9, 2015
Turin, Italy
internazionale
scientifica
Implantable cardioverter defibrillator; Sensitivity; Automatic Testing
no
4 Contributo in Atti di Convegno (Proceeding)::4.1 Contributo in Atti di convegno
Spano, IVAN LUIGI; Sulis, Sara; Serpi, Alessandro; Marongiu, Ignazio; Gatto, Gianluca; Nissardi, V.
273
6
4.1 Contributo in Atti di convegno
reserved
info:eu-repo/semantics/conferencePaper
Files in This Item:
File Size Format  
P51#2015#MEMEA2015#IEEE_Paper.pdf

Solo gestori archivio

Description: An Automatic Sensing Test Procedure for Implantable Cardioverter Defibrillator
Type: versione editoriale
Size 2.21 MB
Format Adobe PDF
2.21 MB Adobe PDF & nbsp; View / Open   Request a copy
P51#2015#MEMEA2015#PostPrint.pdf

Solo gestori archivio

Description: An Automatic Sensing Test Procedure for Implantable Cardioverter Defibrillator (Post Print)
Type: versione post-print
Size 1.77 MB
Format Adobe PDF
1.77 MB Adobe PDF & nbsp; View / Open   Request a copy

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Questionnaire and social

Share on:
Impostazioni cookie